28
Operatingandcommissioning
Type 8228
9.7 calibrating the zero point of
conductivity
→
From the Process level, access the Configuration level.
→
Access the menu "Calib".
→
Got to the function "Zero Calib.".
→
Confirm the function "Zero Calib.".
Calib
Sensor
Probe
Zero Calib.
Calibrate Zero
Point?
Save modified
data?
Processing
The device automati-
cally calibrates the zero
point of conductivity in
less than 1 second.
→
Put the cleaned and dried conductivity sensor in contact with
the ambient air.
→
Choose “Yes”.
→
Save or not the calibration result by choosing "Yes" or "No".
9.8 entering the correction factor of the
fitting used
The correction factor depends on the shape, the material and the
diameter of the fitting used. The following table gives the correction
factors of the fittings S020.
→
From the Process level, access the Configuration level.
→
Access the menu "Calib".
→
Go to the function "K-fitting".
Calib
Sensor
Probe
K-fitting
→
Confirm the function "K-fitting".
→
Enter the correction factor of the fitting used.See Table 1.