Maintenance and troubleshooting
Type 8098
19.11.9
Message "Sound conductivity out of range"
Message
Sound conductivity out of range
Symbol displayed
Depends on the device status the event is associated to.
Possible cause
There are gas bubbles or solid particles in the liquid.
The related event must be enabled. See chpt.
toring special events in the process, on the sensor or on the electronics.
What to do?
→
Search for malfunctions in the process.
→
Make sure the liquid has no gas bubbles and no solid particles.
19.11.10
Message "AO1 open loop" or "AO3 open loop"
Message
AO1 open loop
or
AO2 open loop
Symbol displayed
Depends on the device status the event is associated to.
Possible cause
There is a connection problem on the related output.
The current measured in the current loop is too low compared to the
expected output current.
The related event must be enabled. See chpt.
toring special events in the process, on the sensor or on the electronics.
What to do?
→
Make sure the wiring of the related output is correct.
19.11.11
Message "AO1 Diag error" or "AO3 Diag error"
Message
AO1 Diag error
or
AO2 Diag error
Symbol displayed
Depends on the device status the event is associated to.
Possible cause
There is a connection problem on the related output or a high resistance is
detected in the loop.
The related event must be enabled. See chpt.
toring special events in the process, on the sensor or on the electronics.
What to do?
→
Make sure all the cables are correctly connected.
→
If the related analogue output is not used, disable it. See chpt. 16.4 Dis-