Outputs - Maintenance
Type 8098
18.2
Checking the correct operation of an analogue
output
To check the correct operation of an analogue output, do the following:
→
Connect a multimeter to the analogue output you have adjusted.
→
Select the device in the navigation area.
→
Outputs
→
Maintenance
→
Output 1: analog
or
Output 3: analog
→
Test
→
Enter the current value to be tested
The device generates the entered current value on the selected analogue output.
→
Check the value on the multimeter.
→
Finish
18.3
Resetting the calibration data of an analogue output
to the default values
To reset the calibration data of an analogue output to its default values, do the following:
→
Select the device in the navigation area.
→
Outputs
→
Maintenance
→
Output 1: analog
or
Output 3: analog
→
Reset to default
→
Finish
The calibration data of an analogue output is reset to the default values.