Outputs - Maintenance
Type 8098
18
OUTPUTS - MAINTENANCE
Chapter only valid for variants with outputs.
The settings can be made with the
Installer
user level.
18.1
Calibrating an analogue output
The analogue outputs are calibrated at the factory.
To adjust the analogue output to your equipment, do the following:
→
Connect a multimeter to the analogue output you want to adjust.
→
Select the device in the navigation area.
→
Outputs
→
Maintenance
→
Output 1: analog
or
Output 3: analog
→
Calibration
The current settings are displayed.
The device generates a 4 mA current on the selected analogue output.
→
Enter the current value measured by the multimeter
The device generates a 20 mA current on the selected analogue output.
→
Enter the current value measured by the multimeter
The new settings are displayed.
→
Finish
The analogue output is adjusted.