162
Outputs - Diagnostics
Type 8098
17
OUTPUTS - DIAGNOSTICS
Chapter only valid for variants with outputs.
17.1
Analogue output: reading out the current status and
the values of the current
Any user can read out the following data related to an analogue output:
•
the current status of the analogue output, i.e.
OK
,
Open loop
or
Impedance too high
.
•
the value of the current related to the measured quantity of the process value,
•
the value of the current transmitted on the analogue output.
This data is in read-only mode. To read out some data related to an analogue output, do the following:
→
Select the device in the navigation area.
→
Outputs
→
Diagnostics
→
Select the analogue output
The data related to the analogue output are displayed.
17.2
Digital output: reading out the mode, the current
status and the current value
Any user can read out the following data related to a digital output:
•
the current mode, e.g.
pulse
, of the digital output,
•
the current status of the digital output, i.e.
OK
or
Overload
.
•
the current value of the digital output, e.g. for a pulse output, the number of pulses transmitted on the
output.
This data is in read-only mode. To read out some data related to a digital output, do the following:
→
Select the device in the navigation area.
→
Outputs
→
Diagnostics
→
Select the digital output
The data related to the digital output are displayed.