Document Purpose

Step-by-step instructions with accompanying pictures for donning and doffing SALUS HC are below. Leverage this

document to create and modify a donning and doffing reference to accommodate your standard operating procedures. Please refer to 
the SALUS HC Quick Start Guide or Online Interactive User Manual for more information about your powered air-purifying respirator 
(PAPR) 

www.bullard.com/salusHC/getstarted

1.

Closely inspect the entire PAPR (Powered Air-Purifying Respirator) system paying attention to 
component connection points for wear and damage. If parts are missing or damaged, replace them with
only SALUS HC replacement parts before proceeding. Ensure a battery is installed and has a sufficient 
charge.

2.

Adequate air flow check. With the blower motor OFF, press and hold the power button for ten (10) 
seconds until there is an audible sound. This audible sound indicates that there is adequate airflow to 
create positive pressure inside the hood.

3.

With SALUS HC completely assembled, place the PAPR over your head and onto 
your shoulders. Balance the carriage front to back for a comfortable fit.

4.

Turn the blower motor unit On. Press the button one time to turn the blower motor unit ON. The 

blower unit automatically turns ON at the High-speed setting after approximately 3 - 4 seconds. Press the 
button again to toggle from High-Speed and Low-Speed.

Содержание SALLUS HC

Страница 1: ...age If parts are missing or damaged replace them with only SALUS HC replacement parts before proceeding Ensure a battery is installed and has a sufficient charge 2 Adequate air flow check With the blower motor OFF press and hold the power button for ten 10 seconds until there is an audible sound This audible sound indicates that there is adequate airflow to create positive pressure inside the hood...

Страница 2: ......

Страница 3: ...5 Don the hood starting with your chin and then pull over your head 6 Connect the buckles on the body straps Adjust body straps to ensure a comfortable and stable fit ...

Страница 4: ... motor must not be covered by the isolation gown to allow the respirator to actively draw air through the filter and into the hood or breathing zone Air flow may be reduced and trigger the low airflow alarm if the isolation gown is covering the air inlet or filter of the respirator ...

Страница 5: ... remove it following your hospitals protocols 2 Disconnect both body straps and allow to hang freely at your side 3 Slide the ball of the carriage release hook down and through the opening on the left shoulder of the carriage ...

Страница 6: ...shoulder carriage and entirety of the PAPR system to the right side of your torso to prevent the potential contamination of your breathing zone 5 With the breathing tube still connected doff the hood back to front ...

Страница 7: ...ce on Personal Protective Equipment PPE To Be Used By Healthcare Workers during Management of Patients with Confirmed Ebola or Persons under Investigation PUIs for Ebola who are Clinically Unstable or Have Bleeding Vomiting or Diarrhea in U S Hospitals Including Procedures for Donning and Doffing PPE https www cdc gov vhf ebola healthcare us ppe guidance html ...

Отзывы: