COHP CO Monitor 

User Instructions

www.bullard.com

This unit is designed to monitor breathing air for carbon monoxide contamination from a 
compressed air source. The air flows through the sampling hose and into the monitor where a small 
amount (~0.5 lpm) is diverted through a preset internal regulator and flow restrictor. This delivers 
a continuous flow of sample air to the sensor chamber. The monitor will analyze the air and display 
the CO concentration on the face of the monitor in parts-per million (ppm). 
The Green LED located on the outside of the COHP will illuminate during normal operation when 
the CO level is below the CO alarm point (10 ppm US and 5 ppm International). If the CO level rises 
above the alarm set point, the Red LED on the outside of the COHP will illuminate and an audible 
alarm will sound. 
The unit is available housed a protective polycarbonate case that can be carried or mounted on the 
wall. All system lights and alarms are located on the outside of the polycarbonate case. Grade-D air 
must be supplied to the system and must be free of oil and water to prevent contamination of the 
sensor. The monitor can operate on 115 VAC or 12 volts DC.

Function Modes

The knob on the front of the COM10 or COM5 monitor sets the functions for the CO monitor and can 
be switched to RUN, TEST and CAL.

RUN

 mode is the operation/detection mode. The monitor must be in the RUN mode to monitor the 

supplied breathing air. Supplied air must be turned on and flowing to the instrument while in this 
mode. When the supplied air is off or interrupted while the monitor is in the RUN mode, the low flow 
alarm will sound and LF will be displayed on the monitor until the supplied air is turned on.

TEST

 mode allow for bump testing of the sensor or for silencing of the low flow alarm when 

supplied air is shut off. To perform a bump test, place the monitor in the TEST model and flow 
the calibration gas into the center of the knob.  The reading on the monitor should be 10 

+/–

 1 if 

applying CO gas. The reading on the monitor should be zero or 1 if applying the zero gas. In the test 
mode, the supplied air will not be monitored and there is no need to turn the supplied air off when 
entering this mode. Unit must be zeroed after a bump test is performed. 

CAL

 is the calibration and zero calibration mode. When the monitor is switched to CAL, the supplied 

air will not be flowing to the sensor and there is no need to shut off the supplied air. When in the 
CAL mode, (AC) AutoCal® will be displayed. Pressing the On/Off button once while in this mode 
will change the display to (AO) AutoZero. For calibration and zeroing instruction, see the Calibration 
section.

F1 (fault) 

is indicated on the display when the knob is not positioned securely in one of these 

three modes.

     

indicates that the system requires a re-calibration

     

indicates a low battery

COHP Specs

Case Size:   ...........................................................................................................................17” W x 13”H x 7”D
Weight:  ...........................................................................................................................................................11 lbs
Case:  .........................................................................................................................................................Pelican® 
Monitor Display:  ...................................................................................................................................2 digit LCD
Sensor Type:   ............................................................................................Electrochemical; Carbon Monoxide
Detectable Range:  .................................................................................................................................0-99 ppm
Power:  ..................................................................................... 110/220VAC 50/60Hz or 12 volts DC backup
Operating Temperature: ................................................................................. -4º to 120ºF (- 20º to 48.89ºC)
Operating Pressure: .......................................................................................................................5 psi - 120 psi
Calibration:   ...................................... Manual CO and Zero - Auto-Cal - no manual adjustments required
Humidity Range:  ........................................................................................................................................0-100%
Alarm Setting: .....................................................................................................10 ppm (5 ppm International)
Shielding: .................................................................................................Internal RFI/EMI coating on Monitor
Normal Operation:   .....................................................................................................................Green LEDLight
Alarm Signals:   
 

High CO  ...................................................................................................................Red LED light

 

High CO  .....................................................................................................Audible alarm (90dB)

 

Low Battery  .............................................................................................Amber light (Blinking)

 

Low Battery ...............................................................................................Audible alarm (chirp)

 

Low Flow ....................................................................................................Audible alarm (90dB)

Warranty: ....................................................................................................Two years on sensor and monitor

CAB15 (15 cfm)

Description 

Part Number 

Carbon Monoxide Monitor 

COM10 or COM5

Remote Alarm Jack 
Inlet Fitting- Coulder Coupler
Pressure Relief Valve  

41PRV

Sampling Connector Kit 

COHPKIT

  Incudes:  
  3/8” x 1/4” Adaptor 
  Colder Coupler 
  Tee fitting 
  Reinforced Sampling tube with Colder plugs – 5 ft. length

CO Monitor

Pressure Relief

Valve

Inlet port

Remote 
Alarm 
Jack

Содержание COHP

Страница 1: ...ibrationgasintothecenteroftheknob Thereadingonthemonitorshouldbe10 1if applyingCOgas Thereadingonthemonitorshouldbezeroor1ifapplyingthezerogas Inthetest mode thesuppliedairwillnotbemonitoredandthereisnoneedtoturnthesuppliedairoffwhen enteringthismode Unitmustbezeroedafterabumptestisperformed CAListhecalibrationandzerocalibrationmode WhenthemonitorisswitchedtoCAL thesupplied airwillnotbeflowingtoth...

Страница 2: ...mbienttemperatureschange allowtheunittostabilizefor15 20minutespriortoZeroPointadjustmentandcalibration Zero Point Adjustment ZerocalibrationshouldbeperformedbeforecalibratingthemonitorwithCOgas Bullard recommendsusingimpurityfreetestgastore zeroandwheneverthezero pointhasdrifted Zero Adjustment Procedure RemovethefrontcoveroftheCOMonitor PushtheONbutton Theunitwillbeepandbegina60secondcountdown R...

Страница 3: ...emovingthefrontofthehousingfromtheCOM5orCOM10 monitor Togainaccesstothesensor removethefourscrewsonthefrontoftheCOmonitoronly After removingthefourscrews gentlyprythefrontofthehousingofffromthebottomofthehousing warning Removing the front of the housing by prying at the top will result in damage to the housing This damage is not covered under warranty Gentlypullouttheoldsensor Removetheshortingcli...

Страница 4: ...on 7 If the instrument will not calibrate replace the sensor and repeat both the zero point adjustment and the 10 ppm CO calibration Monitor will not successfully complete a zero point adjustment Wait 5 minutes and repeat steps 1 4 above Monitor displays a 0 reading or rc Wait 5 minutes and repeat steps 1 6 above Monitor zeroes during the zero point adjustment and the reading elevates after return...

Отзывы: