4.1.1 Outdoor installation
The sample gas pumps were not specifically designed for outdoor setup. The operating and environmental conditions are cru-
cial for the required types of protection and any additional measures required, such as:
– adequate protection from the weather
– Adjusting the maintenance intervals (e.g. cleaning and replacing wear parts)
Use suitable measures and regular inspections to prevent damage to the equipment from e.g.:
– Corrosion
– Sunlight (temperature peaks and damage from
UV rays)
– Moisture from condensation (e.g. due to rapid temperature changes or downtimes)
– Icing
– Insects and microbes
– other animals, e.g. martens, etc.
Please remember that all technical operating parameters of the equipment must also be met with outdoor installation. Specific-
ally:
– Maximum or minimum operating temperatures
– Degree of protection
4.2 Installation
CAUTION
Damage to the device
Protect the device, especially the gas inlets and tubes, against dust, falling parts and ex-
ternal impact.
P2.2 Atex/P2.72 Atex
When installing the P2.2 Atex/P2.72 Atex on mounting plates, use the included mounting bracket and only the included rubber/
metal bumpers. Operation without rubber/metal bumpers is prohibited. These must also be used when installing the pump on
an existing substructure. For the hole pattern in the mounting bracket and the motor foot, please refer to the Technical Data at
the end of the operating and installation instructions.
P2.4 Atex/P2.74 Atex
When installing the P2.4 Atex/P2.74 Atex sample gas pump, please note the assembly drawing
42/025-Z02-02-2
. Before begin-
ning the installation, verify the sample gas pump is complete. You will also require 6 x suitable length M6 bolts and nuts for in-
stallation.
The pump head on all pump types can only be aligned turned by 0° or 180°.
11