Nivotemp NT M-XP
5.9.4 Define switching output to log
Used to select the switching output to be logged. Only one switching output can be logged.
Diagnostic
diA
Diagnostics menu
SJ.ou
Output monitoring
Set Journal Out
out_
Out I - Out_
Selection menu
Options:
out1
to
out
x
Switching output logging.
NOTICE
Values are backed up from volatile to non-volatile memory approx. every three hours.
5.9.5 Delay for storing the Min/Max Liquid Level
A delay time for saving the minimum and maximum liquid level can be set to record reliable values when fluid levels are un-
stable. Here, enter the time span in seconds during which the signal must be continuously present before the liquid level is
logged.
Diagnostic
diA
Diagnostics menu
do.MM
Min/Max Delay
Delay MM
___
0 - 100
Selection menu
Setting range:
0…100 seconds
– Use the key to open the list of values.
– Set the value with the
and
keys and use the key to confirm (e.g. 5 (seconds).
The device will return to the submenu.
5.9.6 Delay for storing the Min/Max Temperature
A delay time for saving the minimum and maximum temperature can be set to record reliable values when temperatures fluc-
tuate. Here, enter the time span in seconds during which the signal must be continuously present before the temperature is
logged.
Diagnostic
diA
Diagnostics menu
dt.MM
Min/Max Delay
Delay MM
___
0 - 100
Selection menu
Setting range:
0…100 seconds
– Use the key to open the list of values.
– Set the value with the
and
keys and use the key to confirm (e.g. 5 (seconds).
The device will return to the submenu.
26
Bühler Technologies GmbH
BE100023 ◦ 06/2021