24
MA112680_39.3
[Original Instructions]
4/2014
Warning Messages
The following Warning Messages will appear on the LCD Screen:
HOOD OPEN
The hood is not fully closed.
Check to see if any specimens or other debris is on the hood ledge.
ARM LIMIT
The IsoMet 4000 has a safety switch to limit the blade’s forward travel. Use the hand crank to reposition
the blade.
BLADE PINCHED
Indicates that during cutting the blade became pinched by the specimen and the cutting operation has
automatically shut off.
Carefully retract the blade and check the specimen fixture.
EMERGENCY STOP
Indicates the Emergency Stop button has been pressed.
1. Examine the machine for problems.
2. Rotate the Emergency Stop button counter-clock wise until it pops out
3. Press the CUTTING CYCLE button to continue operation.