8
MA112680_39.3
[Original Instructions]
4/2014
IsoMet 4000 Controls and Functions
Before operating the IsoMet
4000 lightly oil all exposed metal parts. During operation the coolant will
maintain a protective film of oil on all exposed metal.
When the IsoMet 4000 is not in use raise the hood. This will minimize possible corrosion.
A glass-reinforced, splash-proof plastic hood encloses the entire cutting area to prevent loss of coolant
and decrease noise. When raised, the counter-balanced hood will remain in the open position, activating
a safety-lock switch disabling the controls for the cutting motor.
1. Activate power to the IsoMet
4000. Flip the power switch on the back of the machine to the ON
position.
2. On the front control panel press the ON button.
•
The front panel LCD will light up and scroll through the display screen.
•
Raise the hood to allow access to the cutting bay.
Front Panel Controls
POWER
OFF
ON
R
IsoMet 4000
LINEAR PRECISION SAW
1
2
3
4
5
6
EMERGENCY
STOP
The front panel consists of six (6) dedicated buttons, five (5) software/multi-function softkey
buttons, an Emergency Stop button, and a large Liquid Crystal Display (LCD) with back lighting.
All buttons have tactile feedback.
ON
Power ON
The Power ON button will activate the IsoMet system.
When first turned ON, the LCD screen will scroll through the Buehler name, logo,
type of machine, and latest firmware revision.
After 10 seconds the screen will change to display the BLADE SPEED, FEED RATE,
and DISTANCE REMAINING parameter information as well as the softkey button
commands for the L1 screen.
OFF
Power OFF
The Power OFF button will deactivate the IsoMet system. The Power
OFF button will store all previous set parameters before powering off.