BUCKLEYS 

 

BATHYCORROMETER® 

 

INSTRUCTION 

 

MANUAL 

 
 
 

READ THIS MANUAL BEFORE USING THE EQUIPMENT. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Buckleys (UVRAL) Ltd. 

 

Beta Works, Range Road, Hythe, Kent. United Kingdom. CT21 6HG. 

 

Tel:- +44 (0) 1303 260127/8   Fax:- +44 (0) 1303 262115   

 

E-Mail:- [email protected]   web site:- www.buckleys.co.uk 

Содержание BATHYCORROMETER

Страница 1: ...TION MANUAL READ THIS MANUAL BEFORE USING THE EQUIPMENT Buckleys UVRAL Ltd Beta Works Range Road Hythe Kent United Kingdom CT21 6HG Tel 44 0 1303 260127 8 Fax 44 0 1303 262115 E Mail sales buckleys co...

Страница 2: ...nical Data 3 3 1 BCM Unit 3 3 2 Charger 3 4 0 Operation 4 4 1 Charger 4 4 2 BCM Unit 4 5 0 Maintenance 5 6 0 Reference Electrode 5 Replacement 7 0 Safety 6 7 1 Electromagnetic Compatibility 6 8 0 Cali...

Страница 3: ...The corrosion potential is detected by a Silver Silver Chloride Ag AgCl reference electrode of extremely robust and proven construction mounted in the protective nose cone and sited only 5cm from the...

Страница 4: ...3 1 BCM UNIT Operating depth to Up to 300m Pressure tested to 600psi 42bar Display 3 digit LCD backlit 0 001 to 1 999V DC Accuracy 0 05 typical 1 count Input impedance 10MW Operating temperature rang...

Страница 5: ...after use by giving a charge of 12 minutes for each hours use 6 There is no indication on the charger or the BCM that the battery is charged If the charger is turned off it will not discharge the BCM...

Страница 6: ...reference electrode at regular intervals If the unit has been used in a dirty environment the red nose cone should be flushed with clean sea water to remove any contamination which would shorten the l...

Страница 7: ...cone Grease and replace the white Delrin Probe Cone onto the Tie Rod 12 Replace the stainless steel Probe Tip if fitted 7 0 Safety All diving operations which involve the use of the Bathycorrometer mu...

Страница 8: ...ar the O ring on the Calomel Reference Electrode with a little Silicone Grease 3 Screw the Calomel Reference Electrode onto the threaded Tie Rod ensuring a tight connection and correct seating of the...

Страница 9: ...gnificant variation occurs Differences in the order of 10mV or so between readings are quite possible and will mainly be caused by variations in water salinity at different locations or due to changes...

Страница 10: ...ase ready for storage 3 The BCM can be stored with the battery in any state of charge 4 Place the BCM Charger probes probe spanner Silicone grease and this Instruction Manual into the carrying case 5...

Страница 11: ..._______________ Serial No ______________ Accuracy Date Calibrated ______________ Electronics Tested against Test Jig No ____________ Calibrated on ____________ BCM Calibrated against the best of 3 No...

Страница 12: ..._____________ has been manufactured in conformity with the following standards and specifications Safety IEC 1010 1 1990 and amendments 1 1993 and 2 1995 EMC EN 55081 2 1994 Generic Emissions Standard...