Marine Survey Kit 

Operating Instructions 

English Version 

 

Please read carefully before use 

 

Содержание 6004-0093

Страница 1: ...Marine Survey Kit Operating Instructions English Version Please read carefully before use...

Страница 2: ...Contents Description 3 Technical data 3 Unpacking 3 Operation 4 Surveying 4 Storage 5 Calibration 6 Disposal information 6 Contact Details 7 Distributor details 7...

Страница 3: ...cture to the accurate multimeter In use the multimeter is connected to the structure terminal and the cable spool terminal The half cell is then lowered into the sea at the desired position Turning on...

Страница 4: ...are connected correctly x Decide where on the structure the readings are to be taken xi For consistency the reading should be taken in the same place at the same state of the tide if applicable and a...

Страница 5: ...lied with the multimeter Discard the old batteries iv Pack in a cardboard box along with this manual and the multimeter manual Do not put the half cell in a plastic bag Re commissioning i Fit new batt...

Страница 6: ...he multimeter and the K SAT reference cell to the negative terminal With the half cell in a bucket of clean seawater place the tip of the K series reference cell into the water The multimeter should r...