39
BÜCHI
Vacuum Controller
V-800/V-805
11 The ventilation valve
Automatic ventilation
With standard configuration ex factory, when the controller
stops, the connected apparatus is ventilated immediately and
the valve remains open even after the apparatus has been
turned off. As a result, accidental excess pressure situations
are avoided.
However, you can also configure the controller so that hen
the system stops the apparatus is not ventilated. In this case,
you must puss on the Stop key again to ventilate the apparatus.
Attention: Excess pressure situations can arise when the se-
cond operating mode is used. Thus, always remove the
evaporating flask immediately and drain receiving flask!
Configuration possibility: ventilation ON/OFF
12 Remote control
(optional accessory)
Plug remote control to connection RC. Now, remote control
is ready and when the Vacuum Controller is turned on, you
can use it to control all controller functions. The function keys
F1 to F4 correspond to the menu driven keys (visual contact
with Vacuum Controller is precondition).
You can also operate the apparatus even if the remote control
is connected to the Vacuum Controller.
11 Ventilation valve/12 Remote control
Remote control (optional accessory)