6 Functions V-805
6
Functions V-805
6.1
Overview of program structure V-805
6.2
General functions
6.2.1
Manual distillation
The set point of system pressure is input. When the distillation
starts, the system pressure is reduced to the input value and
kept constant. An input value stays in the memory even when
the apparatus is turned off. Distillation must be stopped
manually.
6.2.2
Distillation with pressure gradients
Programming of a pressure profile from up to 20 segments.
Each segment is defined by start pressure, end pressure and
its duration. At the end of the entire pressure profile, the
distillation is stopped. If so configured, rotation of the
evaporating flask is stopped and the flask is lifted from the
bath.