Syncore
Platform
33
5.14 Trouble-shooting: The single platform
Trouble-shooting: the single platform
Observation
Cause
Correction
Display does not light
No power
Connect the unit to the power system.
Check fuses. Check the power system
Display lights only in part
Display unit defective
Contact BÜCHI service agent
Unit does not heat up
Heater not turned on
Press „START“
Unit does not heat up
Heater not turned on
Temperature set-point too low; Setpoint
temperature higher than room temperature.
Unit does not heat up
Heater defective
Contact BÜCHI service agent
Unit does not shake
Drive not activated
Turn speed control clockwise
Unit does not shake
Drive blocked by the adjustment
wrench
Remove the adjustment wrench
Unit does not shake
Drive belt torn
Contact BÜCHI service agent
Unit does not shake
Drive defective
Contact BÜCHI service agent
Vacuum cover is not heated
The 24 V connection socket has not
been connected correctly
Check the plug on the vacuum connection
Vacuum cover is not heatedt
Setpoint temperature too low
Raise the setpoint temperature
Vacuum cover is not heated
Heater defective
Contact BÜCHI service agent
Unit cannot be programmed.
Program mode has not been
activated
Press the „PROGRAM“ button until 1 appears in
the Step display.
Buttons do not react to pressure
Defective contact
Contact BÜCHI service agent
Error message in the display
Mechanical or electronic defect
Contact BÜCHI service agent, refer to Chapter
„Maintenance: Error messages“
Unit shakes but makes noise
when doing so
Vibration elements defective
Inspect vibration elements visually and contact
BÜCHI service agent
Strong vibrations
Settings for unbalance not optimum
Readjust unbalance compensation
Strong vibrations
Rotational speed not properly
adjusted
Reduce speed
5 Operation