4.4
Function test for drive and speed indica-
tor
There should be no objects of any kind resting upon the
base plate or between the base-plate and the housing of
the equipment.
• Set a speed of 60 RPM by turning the speed controller
to the right.
• The display unit
should indicate “60” and the base
plate of the basic unit should move slowly without any
pronounced vibration.
• Following completion of the function test, reset the speed
to 0 RPM
4.5
Function test heating and time
There should be no objects of any kind resting upon the
base plate or between the base-plate and the housing of
the equipment.
• Press once on the „UP“ button
on the temperature
con trol unit. The display
shifts from the current tempe-
ra ture to the set-point temperature.
• Press the „UP“ button repeatedly to set a new set-point
for temperature of +40°C.
• After 3 seconds, the unit takes over this new set-point.
• Press the „START“ button repeatedly. The base plate is
heated to the set-point temperature. A dot in the tempe-
rature display
shows that the heater has been turned
ON.
• The time display
shows how much time has passed
since starting.
If not, contact a local BUCHI service agent.
• Turn the heater OFF after the functional test by pressing
once on the “STOP” button.
• To check the temperature of the heating plate, insert an
external thermometer (Ø 4 mm) into the reference socket.
4 Putting into Operation
Figure 9: Controlling the basic functions from the control panel