6 | Software
Büchi Labortechnik AG
32/62
Operation Manual Sepiatec SFC-250
6.4.2
Setting gradient mode
Navigation path:
➔
PARAMETER
u
Tap the
[GRADIENT]
button.
ð
A new window opens displaying the gradient
table.
u
Tap into the gradient table.
ð
A new window opens to edit the gradient
table.
u
Tap on the field on the control panel with a
blue or white background to enter time,
modifier and flow values.
u
Enter the value.
u
Tap the
[OK]
button to confirm.
ð
The value is saved.
ð
The dialog box closes.
NOTICE! The necessary volume of CO₂ is calcu-
lated automatically.
u
Tap the
[ADD]
button to insert a row at the
bottom of the table.
u
Select a row and tap the
[DELETE]
button to
remove a row.
NOTICE! The course of the gradient is shown in
the upper left part of the gradient window.
u
Tap the
[OK]
button to confirm and return to
parameter menu.
u
Tap the
[CANCEL]
button to discard the changes
and return to parameter menu.
ð
The programmed gradient is displayed.