Büchi Labortechnik AG
Software | 6
Operation Manual Sepiatec SFC-250
29/62
Navigation path:
➔
MANUAL
u
Tap the
[START PUMP]
button.
ð
A new window opens.
u
Tap the
[CHANGE PARAMETERS]
button.
ð
A new window opens to change the flow and
the modifier percentage.
u
Enter the value.
u
Tap the
[OK]
button to confirm.
ð
The value is saved.
ð
The dialog box closes.
u
Tap the
[INJECTION]
button.
u
Enter the value.
u
Tap the
[OK]
button to confirm.
ð
The value is saved.
ð
The dialog box closes.
u
Tap the
[END INJECTION]
button to stop the
injection.
u
Tap the
[STOP PUMP]
to stop the pump.
NOTICE! An automatic run can be started, while the pump is still running but the
injection has to be finished.
The chromatogram can then be called up in the
COLLECTION
menu. The frac-
tionation parameters can be adjusted for automated further separations.
6.3.3
Setting detector signal to zero
Navigation path:
➔
MANUAL
➔
[SERVICE]
u
Tap the
[AUTO ZERO]
button.
ð
The detector signal is set to zero.