6 | Software
Büchi Labortechnik AG
28/62
Operation Manual Sepiatec SFC-250
No.
Name
Description
3
[SERVICE]
button
To open the service settings of the system.
Parameters, sensor signals and valve posi-
tions of the recycling module are displayed.
4
[AUTO ZERO]
button
To set the detector signal to zero.
5
[START PUMP]
button
To start the pump.
[STOP PUMP]
button
To stop the pump.
6
[INJECTION]
button
To start the injection after starting the pump.
[END INJECTION]
button
To end injection.
7
CO
2
VALVE (grey)
Valve is closed
CO
2
VALVE (green)
Valve is open
CO
2
VALVE (red)
The CO
2
inlet pressure is too high, the system
cannot be started.
CO
2
VALVE (yellow/orange
blink)
The CO
2
inlet pressure is too low. The pumps
can be manual started for a flush. Starting an
automated run is not possible
8
[LAMP ON]
button
See RUN METHOD tab.
9
[TEMPERATURE]
button
10
[BACK PRESSURE]
button
6.3.1
Rinsing the tubing and syringe
Navigation path:
➔
MANUAL
u
Tap the
[FILL SYRINGE]
button to rinse the
tubing to the syringe.
ð
The display shows a dialog box with a numeric
input box.
u
Enter the value.
u
Tap the
[OK]
button to confirm.
ð
The value is saved.
ð
The dialog box closes.
6.3.2
Starting the pump
NOTE
Before or during an isocratic method run in after starting the manual mode. By
tapping the
[STOP PUMP]
button a new window opens and allows you to change
the parameters or to stop the pump.