Parameter table
1
2
3
4
5
6
7
No.
Name
Description
1
Fraction Inlet Pressure
The pressure at the input of gas-liquid separa-
tors
2
Recycling Inlet Pressure
The pressure from the CO
2
supply that goes
into the recycling module
3
Recycling Outlet Pressure
The pressure of the recycled CO
2
4
Waste Bottle Sensor
The sensor is set to detect liquid
[ON]
or not
[OFF]
5
Waste Bottle Sensor Count
This value indicates how often the sensor has
detected liquid. This value is an indication
how stable the recycling and the gas-liquid
separators are working. A low value means
stable conditions. A high value means a lot of
liquid is entrained. This value is set to zero at
every start in the Run Method window.
6
Flush Fraction
This value indicates whether the fraction
valve out is closed OFF or open ON.
7
Fraction Open Time
This value indicates that fractionation is in
progress and how long the fraction valve in is
still open.
Functional description
If the Recycling Inlet Pressure and Recycling Outlet Pressure are approximately
equal, the CO
2
pump is supplied with recycled CO
2
.