5 Putting into operation
30
K-415 Operation Manual, Version B
5 .7
Hose connections
NOTE
Use clips and cable ties where necessary to secure the hoses in place.
5 .7 .1
Hose connections of the DuoScrub
The DuoScrub consists of the neutralization and the adsorption vessel.
To establish the hose connections, proceed as follows:
• Connect the adsorption vessel
c
to the bypass valve
d
using the PTFE hose and secure it with
the GL 14 hose connectors at both ends.
• Connect the emission source via a Woulff bottle to the neutralization vessel
b
and secure all
connections with hose clamps.
3
2
1
4
5
a
Neutralization stage
b
Inlet for fumes from the emission source
c
Adsorption vessel
d
Bypass valve
e
Silencer
Fig. 5.11: Hose connections of the DuoScrub