Büchi Labortechnik AG
Installation | 5
Operation Manual Dehumidifier S-396 Nano
15/24
5
Installation
5.1
Before installation
NOTICE
Instrument damaged if switched on too early.
After transporting, wait twelve hours before switching on the instrument. The fluid
in the cooling system requires twelve hours to collect in the refrigerant compressor.
5.2
Establishing electrical connections
NOTICE
Risk of instrument damage because of not suitable power supply cables.
Not suitable power supply cables can cause bad performance or an instrument
damage
u
Use only BUCHI power supply cables.
Precondition:
R
The electrical installation is as specified on the
type plate.
R
The electrical installation is equipped with a
proper grounding system.
R
The electrical installation is equipped with
suitable fuses and electrical safety features.
R
The installation site is as specified in the
technical date. See Chapter 3.5 "Technical
data", page 11
u
Connect the power supply cable to the
connection on the instrument. See Chapter 3.2
"Configuration", page 9
u
Connect the mains plug to an own mains outlet
socket.
5.3
Installations for a spray drying mode
For installations for a spray drying mode, see separate installation manuals.
"Nano Spray Dryer B-90 HP in closed mode"
"Nano Spray Dryer B-90 HP in open mode with gas supply"
"Nano Spray Dryer B-90 HP in open mode"