5 | Installation
BÜCHI Labortechnik AG
34/116
Operation Manual Rotavapor® R-300
u
Remove glass stopcock.
u
Fit the tapered holder (2) over the upper end of the foam sensor (1).
u
Slide the clamp nut (3) onto the opening for the glass stopcock.
u
Feed the foam sensor into the opening for the glass stopcock and through the va-
por duct into the evaporating flask (6) of the Rotavapor R-300 and fix in place with
the clamp nut (3). The tip of the foam sensor should be positioned approx. 4 to 5
cm below the ground neck of the evaporating flask.
u
Align foam sensor centrally and fix in place with the clamp screw (4).
u
Connect the communication cable from the foam sensor to the VacuBox. Plug the
connector into the socket marked "FS".
5.12
Connecting vacuum and coolant tubing
1
2
3
4
Fig. 15: Vacuum and coolant tubing connections on the cooling condenser of the R-300
1
Cooling condenser inlet
3
Cooling condenser outlet
2
Connection for vapor temperature
sensor
4
Vacuum connection
u
Connect coolant tubes to the two coolant connections (1) and (3) using GL14 cap
nuts.
u
Connect the vacuum tubing to the vacuum connection (4) using a GL14 cap nut.
u
Option 1: connect the vapor temperature sensor to connection (2) using a GL14
cap nut.
u
Option 2: connect the vapor temperature sensor of the AutoDest sensor to con-
nection (2) using a GL14 cap nut. See Chapter 5.10 "Fitting AutoDest sensor and
vapor temperature sensor (optional accessory)", page 31.