Operation Manual Rotavapor® R-300
27/116
1
3
2
Fig. 9: Handle of Rotavapor R-300
1
Cover
3
Fixing screw for cover
2
Communciation cable
u
Undo the fixing screw (3) on the underneath of the handle and remove the cover
(1) from the top.
u
Pull the communication cable (2) out of the hole in the arm of the Rotavapor
R-300.
u
Position the holder (2) for the interface unit on the Rotavapor arm (6). At the same
time, feed the communication cable through the bottom hole in the holder.
u
Fix the holder to the Rotavapor arm with the screw (4) supplied. When doing so,
pass the screw through the hole in the metal plate (5).
u
Connect the communication cable to the COM port on the rear of the interface
unit.
u
Position the interface unit (3) on the holder. When doing so, make sure the lugs on
the back of the interface unit engage in the holder.
u
Fix the interface unit on the holder from behind using the knurled-head screw (1)
supplied. When doing so, make sure that the communication cable is not trapped.
5.7
Mounting VacuBox on R-300 (optional accessory)
The VacuBox is required if the R-300 is to be operated with the Interface I-300 or
I-300 Pro.