5 Putting into operation
48
R-220 Pro Continuous Operation Manual, Version A
5 13
Heating bath setup for 4 2 kW heating element
The heating element holder is labelled with the heating power data For the 4 2 version only water
must be used as heating transfer medium Because of the enormous heat input there is a risk of
smoke formation and fire when using other heating transfer media than water!
How to set the maximum setpoint temperature
Limiting the maximal temperature set-point to a value below 100 °C can help to eliminate processing
errors, e g when working with heat-sensitive sensitive substances
AERATE
Temp
+
Set Temp
SET
Store setting
X
Keep the "SET" button pressed while switching on
the Rotavapor® to enter the setup menu
X
Set the temperature limit via the turning knob The
set value is displayed as bath temperature value
on the display
X
Press "AERATE" button to store the setting and to
leave the menu