BÜCHI Labortechnik AG
Help with faults | 8
Operation Manual Rotavapor® R-100
45/60
8.2.2
Replacing the fuse
The fuse is on the rear of the heating bath.
u
Allow the heating bath to cool down.
u
Disconnect the power cord.
u
Open the fuse holder (1) with a large screwdriver.
u
Replace the blown fuse with an intact one of the same rating.
u
Screw the fuse holder (1) back in.
u
Reconnect the power cord.