7.5
Cleaning the heating bath
IMPORTANT
Risk of short circuit from fluid penetration into device
u
Do not immerse heating bath or rotary drive unit in water or pour water over them.
u
Only wipe down the housing with a damp cloth.
The inside of the heating bath bowl should be cleaned regularly and at the latest if:
the heating bath is contaminated
limescale deposits start to form
the stainless steel surface of the heating bath starts to rust
u
Unplug the power supply for the heating bath and rotary drive unit.
u
Allow the heating bath to cool down then empty it.
u
Remove small amounts of limescale in the bowl of the heating bath using non-
abrasive cleaners (e.g. household cleaner and washing-up sponge).
u
Use acetic acid to dissolve stubborn limescale deposits. Afterwards thoroughly
rinse out the bath.
7.6
Removing solvent accumulations
Before any extended period in which the device is not used (e.g. overnight) all fluids
must be removed from the device and the system dried.
u
Fit a clean and dry evaporating flask and receiving flask (see Chapter 6.1.2 "Fitting
the evaporating flask", page 28 and Chapter 6.1.4 "Fitting the receiving flask",
page 30).
u
Make sure that the glass stopcock is closed.
u
Connect the vacuum pump and evacuate the R-100 as much as possible.
u
Allow the vacuum pump to run for a further 2 to 3 minutes.
u
Vent the Rotavapor.
u
Check that all solvent accumulations have been removed.
u
Dispose of solvent residues in accordance with the specifications of the applicable
safety data sheet.