11594031A en
NIRMaster™
08.2018
BÜCHI Labortechnik AG
Meierseggstrasse 40
Postfach
CH-9230 Flawil 1
E-Mail: [email protected]
Connecting a computer to a NIRMaster™ with no embedded PC and finalizing installation.
To initialize a NIRMaster™ and connect to a computer the basic steps are as follows:
u
Connect the instrument to the computer using the ethernet cable supplied.
u
Connect the mains cable to the instrument.
u
Undo the transportation lock.
u
Switch on the instrument, by holding in the power switch on the left hand side.
u
Start the computer and install the NIRWare software.
u
Set up the network settings to allow communication between computer and NIRMaster™.
u
Operate the NIRMaster™ as normal.
Connect the NIRMaster™ to the PC using the supplied ethernet cable
u
Remove the cover from the top of the Instrument with
the key provided. (Two fixings on the left hand side,
two fixings on the right hand side)
u
Route the ethernet cable through the cable gland in the lower chassis.
u
Connect the cable to the ethernet socket on the rear
of the cream colored optical bench.
Connect the mains cable.
ü
The cover is removed.
u
Route the mains cable through the cable gland in the lower chassis.
u
Connect the cable to the main socket at mounted on
the rear of the instrument lower chassis.
NIRMaster™
Supplement Installation Manual