Büchi Labortechnik AG
Cleaning and servicing | 7
Operation Manual NIRMaster™
35/46
u
If substances still adhere, clean the surfaces with alcohol.
u
Avoid contact with hard objects.
u
Do not drop the external reference.
u
Keep it in a dry and clean place.
7.7
Cleaning the spinner
u
The spinner can be cleaned in a dishwasher.
NOTICE! Not cleaning the spinner
on a regular basis can lead to a reduced force transmission of the inner wheel
and inlay. As a result the sample cup will not rotate.
u
The O-ring of the inner wheel needs replacement in case its mechanical treated.
u
Replacement is recommended by a service person.
7.8
Cleaning the add-ons and the custom sample vessels
u
All BUCHI add-ons can be rinsed and cleaned with disinfectants on alcohol basis.
NOTICE! Spray-disinfection should be used in addition to surface disinfection
with wipes only.
u
The following sample vessels can be cleaned in a dishwasher with standard
detergents:
Petri dishes
Unbreakable cup
Spinner cpl.
XL add-on
High performance cup
7.9
How to reset the system fuses
NOTICE
Device fuses may be triggered occasionally (e.g. due to voltage peaks).
When fuses trigger frequently action needs to be taken.
u
Inform your customer service.
u
Do not try to repair or modify the fuses.