7 | Cleaning and servicing
Büchi Labortechnik AG
34/46
Operation Manual NIRMaster™
u
Take out each filter pad.
u
Insert new filter pads.
u
Reassemble the filter frames.
u
Make sure the ventilation slots of the filter frames show into opposite direction
to avoid a shortcut of the cooling airflow.
7.5
Exchanging the inner filters
1
3
2
u
Lift off the top cover.
ð
The filter drawers are accessible below the fans.
u
Push the drawer (1).
ð
The filter holder moves out of its position.
u
Pop out the sandwich filter.
u
For each filter, place a new filter pad between the metal grids (2).
u
Put the filters back in its holders.
u
Push the drawers back into position (3).
u
Close the housing.
7.6
Cleaning the external reference
NOTE
Do not use any cleaning media with abrasive character which might scratch the
surface.
u
For cleaning use abrasion-resistant precision cloths.