7 | Cleaning and servicing
Büchi Labortechnik AG
30/46
Operation Manual NIRMaster™
7
Cleaning and servicing
NOTE
Users may only carry out the servicing and cleaning operations described in this
section.
Any servicing and repair work which involves opening up the casing may only be
carried out by BUCHI service technicians.
7.1
Regular maintenance work
Action
Daily
Weekly
Yearly Additional information
1
Clean the surfaces in direct
contact with the sample af-
ter each measurement in
order to avoid contamina-
tion.
Cleaning the add-ons and
the custom sample vessels
1
After each measurement
Cleaning the external refer-
ence
1
1
1 Check if the filter is clogged
and replace latest every
6'000 h
1 Check if the filter is clogged
and replace latest every
6'000 h
1 When defective, latest ev-
ery 6'000 h
2 When defective, latest ev-
ery 20'000 hours
Extended system service
and testing
2
1 - User; 2 - BUCHI service technicians
7.2
Cleaning the housing
NOTE
Avoid scratches on optical surfaces due to cleaning.
Scratches can influence the measurement precision.
u
Never use any objects for cleaning which might scratch the optical surfaces.
u
Switch off the instrument and all directly connected electronic devices.
u
Clean the housing with a soft, lint free damp cloth with mild soapy water.