6 | Operation
Büchi Labortechnik AG
28/46
Operation Manual NIRMaster™
When the measurement is done (3), the user will be requested to remove the
external reference.
3
Press the ‘Checkmark’ button to finish the procedure.
Frequency of reference measurement
An additional external reference measurement is mandatory:
after an application change
after a lamp exchange and lamp failure
after a service (when the top cover has been reinstalled and locked)
when changing the add-on or measurement setup
at every system start / restart
6.4
How to perform a measurement
Detailed instructions how to perform a measurement can be found in chapter 5
"Creating a qualitative and a quantitative application" of the NIRWare Quick Guide.