Büchi Labortechnik AG
Product description | 3
Operation Manual NIRMaster™
17/46
Component
Material of construction
External reference
Aluminum, glass inlay, PTFE inlay
Transflectance adapter
Stainless steel 1.4435
Detector
InGaAs
Magnets
NdFeB
3.6.4
Installation site
The installation site allows that the power supply can be disconnected at any time
in an emergency.
The installation site has a firm, level and nonslip surface.
The installation site has a fume hood.
The installation site has an own mains outlet socket for the instrument.
The installation site has enough space that cables / tubes can be routed safely.
The installation site has no obstacles (e.g. water taps, drains, etc.).
The installation site is not exposed to external thermal loads, such as direct solar
radiation.
The installation site is not located near vibration-sensitive devices.
The installation site meets the requirements for the connected devices. See
related documentation
The installation site meets the requirements of the safety data sheets for all
solvents and samples used.
The installation site meets the safety requirements. See Chapter 2 "Safety",
page 7
The installation site meets the specifications according to the technical data (e.g.
weight, dimension, etc.). See Chapter 3.6 "Technical data", page 15