15/34
Hard disk space
At least 80 GB free disk space
Use a hard disk suitable for continuous operation.
Data backup
Network or external
hard disk
At least 0.5 GB free disk space
Additional 20 MB per day and sensor
Screen resolution
At least 1280x1024
LAN
At least 1 x 100 Mbit/s LAN
USB 2.0/3.0
At least 1 USB connection per sensor and 1x USB per Data-
Lab I/O box
PCI/PCIe
1 slot for Profibus card
(for Profibus connection)
Software
Word and Microsoft Excel 2003 or later
3.5.7
Installation site (laboratory)
The installation site has a firm, level surface.
The installation site meets the safety requirements. See Chapter 2 "Safety",
page 6
The installation site has enough space that cables can be routed safely.
The installation site has no obstacles (e.g. water taps, drains, etc.).
The installation site has an own mains outlet socket for the instrument.
The installation site is not exposed to external thermal loads, such as direct solar
radiation.
The installation site allows that the power supply can be disconnected at any time
in an emergency.
The installation site meets the specifications according to the technical data (e.g.
weight, dimension, etc.). See Chapter 3.5 "Technical data", page 12
3.5.8
Installation site (production)
The installation site has a firm, level surface.
The installation site meets the safety requirements. See Chapter 2 "Safety",
page 6
The installation site has enough space that cables can be routed safely.
The installation site is not exposed to external thermal loads, such as direct solar
radiation.
The installation site meets the specifications according to the technical data (e.g.
weight, dimension, etc.). See Chapter 3.5 "Technical data", page 12
The installation site meets the specifications according to the installation point.
See Chapter 5.1 "Establishing installation point", page 17
The installation site has a own mains outlet socket for the instrument.
The installation site has a sample removal point at a distance of < 1 m.
The installation site allows a direct product measurement.
The installation site has constant product flow.
The layer thickness of the product to be measured is at least 30 mm.