3 | Product description
NIR-Online GmbH
12/34
Operation Manual X-Sential™
Letter
Option
V
External button
X2
X-Beam 002
(approx. 15 cm measurement distance)
3.5
Technical data
3.5.1
Sensor
Specifications
X-Sential
Dimensions (W x D x H) 200 x 200 x 100 mm
Weight
5 kg
Max. operating pressure 30 bar at flange
Product temperature
(temperature at flange
with water cooling)
-10 °C to +130 °C
Product temperature
(temperature at flange
without water cooling)
-10 °C to + 70 °C
Vibrations
0.2 G at 0.1 - 150 Hz
Wavelength spectrum
NIR range
900 - 1700 nm; 11100 - 5880 cm
-1
Wavelength spectrum
Visible range
350 - 900 nm; 28500 - 11100 cm
-1
Detector
Diode array
Average measurement
time
20 spectra/s
IP Code
IP69 / IPX9K
Type of lamp
Tungsten-halogen dual lamp
Lifetime lamp
18000 h (2 x 9000 h)
Minimum clearance on
all sides
100 mm
Connection Voltage
85 to 264 VAC
Frequency
50/60 Hz
Power consumption
30 W
Temperature stabiliza-
tion
ASDC (Advanced Spectral Drift Control): active tempera-
ture control to ±1°C from set system operating tempera-
ture. Deviations will lead to automatic white reference
measurement to account for spectral drifts.