5 | Installation
NIR-Online GmbH
26/36
Operation Manual Process Analyser X-One
5.9
Establishing the electrical connection to the installation box
WARNING
Death or serious burns by electric current.
u
Have the installation carried out by an electrician or a person with similar expert
knowledge.
u
After installation, check electrical safety.
NOTE
Observe the legal requirements when connecting the instrument to the power sup-
ply.
u
Use additional electrical safety features (e.g., residual-current circuit breakers) to
comply with local laws and regulations.
u
Perform installation in accordance with IEC/EN 60079-14.
The power supply must fulfil the following conditions:
1. Provide the specified mains voltage and frequency.
2. Be designed for the load imposed by the instruments connected.
3. Be equipped with suitable fuses and electrical safety devices.
4. Be equipped with proper grounding.
NOTICE
Risk of property damage and diminished performance due to use of unsuitable
power cables.
The power supply cables supplied with the product precisely match the require-
ments of the instrument. If other power cables that do not meet those requirements
are used, the product may be damaged and/or its performance diminished.
u
Use only the power supply cables supplied with the product or ordered sepa-
rately from the manufacturer.
u
If using any other power supply cables, make sure that they match the specifica-
tions on the type plate.