NIR-Online GmbH
Installation | 5
Operation Manual Process Analyser X-One
23/36
5.6
Sensor installation
DANGER
Use of an unsuitable flange in potentially explosive atmospheres.
The use of an unsuitable flange may cause an explosion.
u
In potentially explosive atmospheres use a double flange.
The following installation positions are possible:
Fig. 7: Fixing sensor with screws
Tools required:
Torque wrench, size Torx T30
Tightening torque: 8.4 Nm +-1
Precondition:
R
Installation point has been established. See Chapter 5.3 "Establishing installation
point", page 20.
R
The fixing points or bolts conform to M6 A2-70 15 mm
u
Fix the sensor to the installation point using the bolts.