7 | Operating Pro control panel
Büchi Labortechnik AG
74/118
Operation Manual Lyovapor™ L-300/L-300 Pro
Setting
Option
Explanation
[Continue]
Yes/No
Yes : the method switches to the
next phase.
No: the phase is ended when the
set levels are reached.
[Message]
Yes/No
The control panel shows or does
not show a message as soon as
the temperature difference test
is passed.
7.9
Performing freeze-drying using a method [Pro control panel]
7.9.1
Preparing the instrument
NOTICE
Several conditioning of instrument in a day is damaging for cooling circuit.
If it is necessary, wait 2 hours between each conditioning.
Time
re-
quired:
approx.
30min
Navigation path
➔
Start
Precondition:
R
u
Navigate to the
[Start]
menu via the navigation path.
u
Tap the function
[[Start conditioning]]
on the function bar.
ð
The temperature in the ice condenser cools down to operating temperature.
ð
The vacuum pump is being brought up to operating temperature.
ð
After completion of the conditioning phase, the status bar shows the status
Unload / Load.
7.9.2
Selecting a method
Navigation path
➔
Method
u
Navigate to the
[Method]
menu via the navigation path.
u
Tap the function
[Method]
on the function bar.
u
Tap the method that you wish to use.
u
Tap the function
[Activate]
on the function bar.
ð
The status bar shows the method activated.