8 Troubleshooting
56
KjelFlex K-360
Operation Manual, Version D
Table 8-1: General malfunctions and their remedy (cont.)
Error number / Error indication
Possible cause
Corrective measure
59
EEPROM, CPU
Contact the Buchi customer service
60
You cannot change the Standard Method!
Store the new parameter under a new
method name
61
You cannot delete the Standard Method!
Do not try to delete the standard method
62
You have changed an existing method.
Overwrite method?
Overwrite the existing method or store the
new parameter under a new method name
63
The method list is full! You have to delete
some methods.
Delete some methods and add new
methods
64
Restart from Watchdog!
No corrective measure needed!
65
You changed between the demo and the
normal mode.
Restart the system to activate the selected
mode
66
Attention: Reposition stop-cock! (Reagent)
Check the correct position of the stop-cock
67
Attention: Reposition stop-cock! (NaOH)
Check the correct position of the stop-cock
8 .2 Diagnostics
The K-360 software enables to carry out some operational, sensor, service and printer test and to
view some hardware information and the operating hours.
Fig. 8.1: Diagnostics
8 .2 .1
Operational Tests
This submenu enables to test the following components:
• Valves
• Pumps
• Stirrer
• Alarm
• Steam Unit Heating
• Titrator Out (Titrator suitable)
• Dispenser Out (Dispenser suitable)
• Battery Monitor