5 Putting into operation
17
K-437
Operation Manual, Version A
5 .3
Hose connections
Fit the EPDM hose to the Scrubber or the water jet pump to exhaust the digestion gases.
NOTE
Make sure that the hose connections are as short as possible.
WARNING
Danger of toxication. When working with a water jet pump, poisonous and caustic gases can escape
into the environment.
Make sure that the suction output is sufficient.
•
ATTENTION
If the Scrubber is used without condensation step, the condensation that accrues must be caught by
means of a Woulff bottle between the Digest System and the Scrubber.