3 Technical data
13
K-437
Operation Manual, Version A
3 .2
Technical data overview
Table 3-6: Technical data K-437
Digest System K-437
Dimensions (W x D x H)
435 x 560 x 780 mm
Connection voltage
200-230 V ± 10 %
Frequency
50/60 Hz
Power consumption
max. 2200 W
Power input
10 A
Weight (net) (including digestion tubes)
35 kg
Installation category
II
Degree of pollution
2
Environmental conditions
Temperature
Altitude
Humidity
for indoor use only
5 - 40 °C
up to 2000 m
maximum relative humidity 80% for temperatures up to
30°C
Temperature range
50 °C to 420 °C (450 °C)
Temperature drift
± 5 °C > 200 °C / ± 2 °C < 200 °C
Table 3-7: Technical data COD Accessory
COD Accessory
Thermal stability of heating block
± 2 °C at temperatures < 200 °C
Heating up period
Up to around 150 °C within 10 minutes
3 .3
Materials used
Table 3-8: Materials used for the K-437
Component
Material designation
Material code
Housing
Sheet steel DC01 + EVZ
DIN 17162
Coating
Polyester/epoxy
PEP 31
Heating block
Aluminium AlMgSi 1
DIN 3.2315
Insulation
Thermal Ceramics 800
X-607
Rack
Sheet steel, stainless
1.4301
Glass components
Borosilicate 3.3
DIN/ISO 3585
Hose
Ethylene/propylene terpolymer rubber
EPDM
Table 3-9: Materials used for the COD Accessory
Component
Material designation
Material code
Rack COD
Steel, stainless
1.4301
Digestion tube holder COD
Steel, stainless
1.4301
Preparation block COD
Aluminium AlMg4.5Mn
H111 En-5083
Rack condenser COD
PMMA
Stand for condenser COD
PP gray