BÜCHI Labortechnik AG
Editing a method | 8
Operation Manual FatExtractor E-500
55/88
Navigation path
➔
u
Navigate to the
[Method]
menu via the navigation path.
u
Tap the name of the method that you wish to edit.
ð
The display shows the view
Method
.
u
Tap the function
[Edit]
on the function bar.
u
Select the action
[Solvent]
with the navigation control.
ð
The display shows a dialog with the selectable solvents.
u
Select the solvent you wish to use with the navigation control.
u
Tap the function
[OK]
on the function bar.
u
Tap the function
[Save]
on the function bar.
ð
The solvent is saved.
ð
The dialog box closes.
8.4
Setting the extraction time of a method
Enter the duration of the extraction step in minutes.
at least 0 minutes
maximum 5940 minutes / 99 hours
Navigation path
➔
u
Navigate to the
[Method]
menu via the navigation path.
u
Tap the name of the method that you wish to edit.
ð
The display shows the view
Method
.
u
Tap the function
[Edit]
on the function bar.
u
Select the action
[Extraction time]
with the navigation control.
u
Turn the navigation control until the required value.
u
Tap the function
[OK]
on the function bar.
u
Tap the function
[Save]
on the function bar.
ð
The value is saved.
ð
The dialog box closes.
8.5
Setting the extraction cycles of a method (Soxhlet Extraction only)
The extraction time and the number of extraction cycles are interdependent.
If the number of extraction cycles is set to zero, the extraction will finish after the
selected time.
If the extraction time is set to zero, the extraction will finish after the number of ex-
traction cycles is reached.
If the extraction time and a number of extraction cycles is set, the extraction will fin-
ish once both settings are fulfilled.
The number of extraction cycles is selectable from 0 to 5940.
Navigation path
➔
u
Navigate to the
[Method]
menu via the navigation path.