BÜCHI Labortechnik AG
Operating the interface pro | 7
Operation Manual FatExtractor E-500
51/88
7.3.3
Configuration menu
On the
[Configuration]
menu, you can enter a variety of settings and retrieve informa-
tion.
Submenu Settings
The submenu
[Settings]
contains system settings for the instrument.
Action
Option
Explanation
[Language]
Choice of display lan-
guage on the interface
The following languages are avail-
able:
English/German/French/Italian/
Spanish/Japanese/Chinese/Russian/
Polish
[Date]
Date input
Enter in sequence: Day, month,
year. Apply the settings by pressing
[OK]
.
[Time]
Time input
Enter in sequence: Minutes, hours.
Apply the settings by pressing
[OK]
.
[Beeper]
Off/On
Setting for audible signal in re-
sponse to input controls.
[Display brightness]
Enter setting
Display brightness level in %: 0 -
100
[Configuration]
Choice of extraction
method
The following configurations are
available:
SOX (Soxhlet Extraction) / HE (Hot
Extraction) / ECE (Economic Contin-
uous Extraction)
[Max. Cycle time]
(Soxhlet Extraction
only)
Enter value
Maximum cycle time in minutes: 0 -
240
[Analyte protection
(option)]
View
Threshold for empty-no beaker
Threshold for full-empty beaker
[Mobile connection
QR code]
View
The control unit shows a QR code
for connecting the Extraction Re-
ports app with the instrument.
[Mobile connection
password]
View
The control panel shows a password
for entry in the Extraction Reports
app.
[Network]
Enter value
The following parameters can be
edited:
Device name/DHCP/MAC address/
System IP address/Subnet mask/
Gateway
[Delete app con-
nection]
Confirmation question
Resets external connections to the
instrument.
[Demo Mode]
On/Off
Simulate an extraction.