7 | Performing an extraction
Büchi Labortechnik AG
70/88
Operation Manual UniversalExtractor E-800
7.5
Ending an extraction
WARNING
Risk of dangerous vapors, hot parts and hot glassware.
A hot instrument can cause building fire, skin burns or death.
u
Wait 15 min before opening the lift.
u
Do not touch hot parts or surfaces.
u
Use pliers to remove the glassware.
Navigation path
➔
Precondition:
R
The
[Home]
menu is showing the status Finished.
R
The instrument temperature is less than 150 °C.
u
Tap the function
[Up]
on the function bar.
ð
The lift moves up.
u
Open the protection shield.
u
Remove the beaker for further processing.
u
u
Switch off the connected recirculating chiller or close the water tap.
u
Empty the solvent tank. See Chapter 8.2 "Removing the tank bottle", page 73
7.6
Shutting down the instrument
Navigation path
➔
Precondition:
R
The extraction process has ended.
u
Switch the On/Off master switch to Off.
7.7
Exporting extraction parameters
NOTE
The number of saved QR Codes is limited.
u
When the internal memory is full, the older QR Codes will be overwritten.
7.7.1
Exporting extraction parameters without cloud connection
Precondition:
R
The App is installed on a mobile device.
R
An extraction is finished.
u
Tap the button
[QR Code]
on the function bar.
ð
The display shows a QR code.