Optimizing basic instrument settings
Analyzer configuration
You can optimize the following by means of configuration:
Interface properties
Signature program properties
Device parameter
See
Configure analyzer
(on page 75).
Reacting to irregularities
If irregularities occur during the analysis, it is important for the user to quickly identify any errors. In
the "Error handling" dialog the user defines how the analyzer is to react to incidents.
Incident
Consequences
During analysis, the computed element
contents exceed the defined tolerances.
If standard tolerance monitoring is enabled, the analysis is
aborted. "Tol" is displayed in the information column of the
particular sample.
A critical error occurs.
The analyzer is stopped after the analysis.
A non-critical error occurs.
If you have set the analyzer to stop after a set number of
non-critical errors and this number has been reached, the
analyzer is stopped after the analysis.
See
Error handling
(on page 328).
Defining custom methods
Depending on what sample substances you want to analyze, it may be necessary to define custom
methods in addition to the default methods.
Methods allow to set certain parameters for sample specific requirements.
See
Defining custom methods
(on page 129).
Sleep/wake-up behavior
For sleep/wake-up behavior, you can specify that conditioning samples in a series are measured
before the first user starts work on the instrument in the morning. See
Optimizing sleep and wake-
up behavior
(on page 132).
Optimizing sample data editing
Facilitating input of samples
In the "Input options" dialog you can define settings that facilitate sample data input in the following
cases:
An identical sample name is used several times.
You need to increment or decrement a figure in the sample names.
One method is used for numerous samples.
You need to input a daily factor for numerous samples.
You need to input a blank value for numerous samples.
See
Input options
(on page 315).
Содержание DuMaster D-480
Страница 1: ...DuMaster D 480 Operation Manual 11593626C en...
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 30: ......
Страница 102: ......
Страница 181: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 8 Maintaining the instrument 181...
Страница 286: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 10 Menu and dialog descriptions 286...
Страница 338: ......
Страница 352: ......
Страница 353: ......