Operating instructions DuMaster D-480
©BÜCHI Labortechnik AG
10 - Menu and dialog descriptions
319
In section "Factor determination":
Label
Element type
Meaning
sequentially
Radiobutton
The daily factor is computed sequentially.
total
Radiobutton
The daily factor is computed as a total.
manual input
Radiobutton
The daily factor can be entered manually, there will be no
calculation of the daily factor.
autom. factor
calculation
Checkbox
calculation of the daily factor starts automatically after the
end of every analysis.
separate upper/lower
calibration range
Checkbox
Determines whether valid factors shall be computed for
both calibration ranges or separate ones for each range.
In section "Blank value determination":
Label
Element type
Meaning
sequentially
Radiobutton
The blank value is computed sequentially.
total
Radiobutton
The blank value is computed as a total.
manual input
Radiobutton
The blank value can be entered manually, there will be no
calculation of the blank value.
autom. blank value
calculation
Checkbox
calculation of the blank value starts automatically after the
end of every analysis.
Remaining elements:
Label
Element type
Meaning
OK
Button
Saves new settings
Cancel
Button
Discards settings not yet saved.
Further information
Further information includes instructions in which the dialog is used and knowledge articles
concerning the dialog:
Determining the daily factor
(on page 142).
Determining blank values
(on page 141).
Specifying the computation method for blank value and daily factor
(on page 121).
LIMS settings (Dialog)
Function
This dialog is used to make various adjustments to a laboratory information and management system
(LIMS). The dialog has the following tabs:
On the "LIMS data" tab you select which data to send to the LIMS and whether they should be
sent automatically after every analysis.
On the "Communication" tab you specify whether and how to transfer data.
If you are using multiple analyzers with one PC, you must specify one instrument identifier on the
"Identifier" tab so that the LIMS data can be allocated to the correct instrument. You can also
define user names.
Note
Note that the LIMS port name must not be identical to the port name between the PC and the
analyzer. This could lead to a system crash.
Содержание DuMaster D-480
Страница 1: ...DuMaster D 480 Operation Manual 11593626C en...
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 30: ......
Страница 102: ......
Страница 181: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 8 Maintaining the instrument 181...
Страница 286: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 10 Menu and dialog descriptions 286...
Страница 338: ......
Страница 352: ......
Страница 353: ......