8 - Maintaining the instrument
202
Press the lower half shells out of the ball valve casing by means of the "knock-out".
Cleaning, assembling and installing the ball valve.
Note
The following section is a continuation of Removing and dismantling the ball valve. All steps
described there must already have been performed.
Process
The section is divided into the following phases:
Cleaning the half shells and ball
Assembling the ball valve
Mounting the ball balve
Preparing the Instrument for measurements
Requirements
Before starting work, the following requirements must be met:
Keep the following tools ready:
Hook wrench
Phillips screwdriver / Allen wrench 3 mm
8 mm open jaw wrench.
Keep acetone ready to clean components.
The ball valve must have been removed.
The instrument must have cooled down and be disconnected from the mains when you install the
ball valve again. See step 1 in Removing and dismantling the ball valve.
Safety instruction
Strictly observe the safety instruction "Hot instrument parts" under
Warning notes during operation
(on page 24).