6 Operation

21

C-660 Operation Manual, Version F

6 .3 .2 

Manual control / manual mode

Overview of the menu navigation

Fig. 6.5: Overview of the menu navigation

Parameters

Table 6-3: Main menu

Main menu

Description

Fraction

Displays the current fraction.

Time or volume

It is possible to collect according to time and volume. This setting is selected in the Parameter 
menu under the Collect point.

Detector

The line appears when a detector has been selected.

Parameter

Jumps into the Parameter Menu.

Содержание C-660

Страница 1: ...Fraction Collector C 660 Operation Manual 096969 en...

Страница 2: ...eserves the right to make changes to the manual as deemed necessary in the light of experience espe cially in respect to structure illustrations and technical detail This manual is copyright Informati...

Страница 3: ...3 Installation 16 5 3 1 Power connection 16 5 3 2 Fitting and ferrule 16 5 3 3 Optional waste diverter valve 16 Read this manual carefully before installing and running your system and note the safet...

Страница 4: ...height of the collector arm 24 6 3 4 Interface RS 232 25 6 3 5 Peak detection 25 7 Maintenance 28 7 1 Cleaning 28 8 Troubleshooting 29 8 1 Malfunctions and their remedy 29 8 2 Customer service 30 9 Sh...

Страница 5: ...s 96961 96965 UV Monitor C 630 Operation Manual numbers 96950 96953 and 96966 UV Photometer C 635 Operation Manual numbers 96957 96960 and 96967 Cartridger C 670 Operation Manual numbers 96945 96949 1...

Страница 6: ...s as a basis for training 2 2 Proper use The Fraction Collector C 660 has been conceived of and constructed as a laboratory instrument Its correct usage is the optimal and reliable collection of indiv...

Страница 7: ...uilt in accordance with current state of the art technology however risks to users property and the environment can arise when the instrument is used care lessly or improperly The manufacturer has det...

Страница 8: ...The operator shall inform the manufacturer without delay of any safety related incidents which might occur during the operation of the instrument Legal regulations such as local state and federal law...

Страница 9: ...r detailed information on the listed products see www buchi com or contact your local dealer 3 1 1 Standard instrument Table 3 1 Standard instrument Product Order number Fraction Collector C 660 44900...

Страница 10: ...French 96971 Italian 96972 3 1 3 Optional accessories Table 3 3 Optional accessories Product Order number Waste valve valve for the Fraction Collector Prevents dripping during the fraction change 511...

Страница 11: ...0 250 mL tubes for rack FC 12 diameter of 39 mm height of 300 mm 44960 i 20 50 mL tubes for rack FC 30 diameter of 25 mm height of 185 mm 44961 j 100 20 mL tubes for rack FC 60 diameter of 16 mm heigh...

Страница 12: ...25 W Frequency 50 60 Hz Installation category II Degree of protection IP20 Pollution degree 2 Environmental conditions Temperature Altitude Humidity for indoor use only 5 40 C up to 2000 m maximum re...

Страница 13: ...signals Display Back lit LCD graphic display 28 64 pixels 58 29 mm Dialog languages en de fr it es jp Interfaces 2 RS 485 system bus 1 RS 232 process data output 1 Mini DIN waste valve 1 Mini DIN TTL...

Страница 14: ...F switch f FC 60 rack with glass tubes g Drip tray Fig 4 1 Example of System 3 4 2 Function principle The fraction collector is characterized by the following features Up to 12 liters 48 250 mL or up...

Страница 15: ...scertained that the instrument is not exposed to aggressive substances such as acids bases and solvents which can have a negative influence on the useful life WARNING The fraction collector is not exp...

Страница 16: ...to firmly seal the tubing end with the ferrule Fig 5 2 Installing fitting and ferrule 5 3 3 Optional waste diverter valve If the flow rate of the pumps exceeds 10 mL min a waste diverter valve is rec...

Страница 17: ...bing guide f so that it pokes out approximately 3 mm Connect the valve to the WASTE VALVE con nection i 9 Fig 5 3 Installating the waste diverter valve 2 3 4 1 8 5 6 7 9 10 Fig 5 4 Control arrangement...

Страница 18: ...Control elements 1 2 3 4 5 a START NEXT button b STOP PAUSE button c Display d Selector switch turn and press knob e ON OFF Power switch Fig 6 1 Control elements of the fraction collector Table 6 1 Fu...

Страница 19: ...llection status PAUSE Pressing the STOP Pause button once The program sequence is interrupted and the BUCHI pump modules are stopped no reset Other brand pumps will not be stopped The program sequence...

Страница 20: ...in the stop mode You set the desired values by turning the selector switch Press the selector switch in order to confirm the new value After 10 seconds without confirmation the cursor jumps back to t...

Страница 21: ...avigation Parameters Table 6 3 Main menu Main menu Description Fraction Displays the current fraction Time or volume It is possible to collect according to time and volume This setting is selected in...

Страница 22: ...e rack in the display and simulates the progression of the collector arm Collecting It is possible to set whether collection should take place in fixed time intervals or according to volume only in co...

Страница 23: ...programs P01 P10 are deleted Table 6 6 Rack definition menu Rack definition menu Description back Jumps back into the Parameter menu Rectangular rack Jumps into Rack Menu 1 Ractangular rack Irregular...

Страница 24: ...erwise the previously defined empty C01 C10 racks C01 C10 can be deleted with a yes no query 6 3 3 Setting the height of the collector arm Setting the height The distance between the collector arm and...

Страница 25: ...n in the Waste column in the Wait Pause and Error states While the arm is moving from one position to the next the new tube number and tube position are entered and the Waste column is not marked The...

Страница 26: ...ially in the case of only slightly separated peaks this procedure allows the multiple extraction of a fraction with sufficient purity for the subsequent analysis A tube change takes place when crossin...

Страница 27: ...n times have considerably flatter flanks and are usually somewhat more agitated such peaks require a more slug gish switching decision The following serves as a rule of thumb Use Range 3 for separatio...

Страница 28: ...or to any maintenance work on the instrument switch off the power supply and remove all sources of flammable vapor ATTENTION Use only genuine consumables and spare parts for any maintenance and repair...

Страница 29: ...it on again if Arm is jammed appears in the display thus enabling the collector arm to reposition itself Please contact the BUCHI customer service if this measure does not help 8 1 Malfunctions and t...

Страница 30: ...ons or application prob lems please contact one of these offices The customer service offers the following Spare part delivery Repairs Technical advice Before calling us Our service personnel will be...

Страница 31: ...nces Clean the instrument and all accessories thoroughly to remove possibly dangerous substances Do not clean dusty parts with compressed air Store the instrument and its accessories at a dry place in...

Страница 32: ...timum system performance and reliability Prior written permission of the manufacturer should be obtained before any modifications are made to the spare parts used 1 3 4 2 Table 10 1 Spare parts Produc...

Страница 33: ...ause harmful interference in which case the user will be required to correct the interference at his own expense Fran ais Cet appareil a t test et s est av r conforme aux limites pr vues pour les appa...

Страница 34: ...11 Declarations and requirements 34 C 660 Operation Manual Version F 11 2 Declaration of conformity...

Страница 35: ......

Страница 36: ...www buchi com cn India BUCHI India Private Ltd IN Mumbai 400 055 T 91 22 667 75400 F 91 22 667 18986 india buchi com www buchi in Indonesia PT BUCHI Indonesia ID Tangerang 15321 T 62 21 537 62 16 F 6...

Отзывы: