3 Technical data
12
C-660 Operation Manual, Version F
3 .2
Materials used
Table 3-4: Materials used
Component
Material designation
Housing
PUR foam
Tubing
FEP
Fitting
POM
Ferrule
ETFE
Drip tray
Steel
Chromatography table
Aluminum
3 .3
Technical data overview
Table 3-5: Technical data
Dimensions ( W×D)
465×470 mm
Height (H)
Adjustable from 370 mm to 460 mm
Weight
14 kg
Connection voltage
100 – 230 VAC ± 15%
Power consumption
max. 25 W
Frequency
50/60 Hz
Installation category
II
Degree of protection
IP20
Pollution degree
2
Environmental conditions
Temperature
Altitude
Humidity
for indoor use only
5 – 40 °C
up to 2000 m
maximum relative humidity 80% for temperatures up to
31 °C, and then linearly decreasing to 50% at 40 °C
Support racks
4× FC60: 4×(60×20 mL tubes)
4× FC30: 4×(30×50 mL tubes)
4× FC12: 4×(12×250 mL tubes)
Syncore racks (R4, R6, R12, R24, R48, R96)
Teach-in-function of 11 user-specific racks
Collection area (W×D)
440 mm×350 mm
Collection modes
Time (interval timer 1 s – 100 h)
Volume (only with BUCHI Sepacore pumps)
Peak threshold (together with a detector)